Piezoelectric micropumps

 

Due to advantages of fluidic microsystems over conventional systems (small size, flexibility, low reagent consumption, low energy consumption, speed of analysis, mobility, low cost) they are becoming increasingly employed in the field of chemistry, molecular biology, pharmacy, medicine, etc. Basic micro fluidic system is composed of micropump which pumps the liquid through the microchannels. For a good, reliable and long term performance of fluidic microsystem, design and fabrication of micropumps are usually the most challenging and critical parts. Therefore we investigated microfluidic phenomena in electromechanical-fluidic systems in order to improve reliability and performance of piezoelectric micropumps.

 

The suitability of micropumps for real-life application is ensured by reliable operation, long-term stability, the ability to pump two-phase medium (bubble tolerance), the ability of self-priming, low probability of cavitation, chemical stability of the materials with respect to the pumping medium and performance.

 

In reciprocating displacement micropumps, the cavitation is associated with high-velocity media initiated by mechanical vibrations of the membrane. To suppress cavitation, one should raise the static pressure or reduce the fluid velocity at cavitation nucleus. Raising the inlet pressure, proper priming procedure, proper geometry and relief of channels, precise valves phase sequence during operation, amplitude / frequency / wave-form of the excitation signal and fluid cooling greatly reduce the risk of cavitation. To design reliable and cavitation free piezoelectric micropumps, cavitation phenomenon was investigated in detail using numerical simulations and experiment analysis.

 

Operation reliability is desired quality of every microfluidic application. If the pump is not bubble tolerant, the presence of gas in the medium might result in a permanent failure of the entire system. If micropump is not self-priming, additional priming equipment and priming procedures are required which aggravate micropump start-up. Micropump bubble tolerance can be improved by increasing the compression ratio, whenever possible.

 

Design & numerical simulations

PDMS elastomer is cast using double-depth molds. See photomasks for mold fabrication on the image below.

We developed a fully coupled 3D electro–fluid–solid mechanics simulation model that enables designing, numerical modeling, virtual (simulated) testing and numerical optimization of micropumps with piezoelectric actuation. Advanced simulation model covers fluid inertial effects, a hyperelastic  model  for  PDMS  and no-slip  boundary  condition  in  fluid-wall  interface.

 

 

 

Opis: Opis: Opis: Opis: Opis: abstract graphicOpis: Opis: Opis: Opis: Opis: C:\Users\zummy\Desktop\verzija\nove slike\Figure4.tifOpis: Opis: Opis: Opis: Opis: Opis: ani10.png

Fabrication

In the Laboratory of Microsensor Structures and Electronics we have developed the entire process technology for micropumps production, so that it can be cost-effective to produce even a small series of dedicated products.

Opis: Opis: Opis: Opis: Opis: 22Opis: Opis: Opis: Opis: Opis: 44Opis: Opis: Opis: Opis: Opis: DSCN7577

Characterisation

For characterization purposes, automatic characterization setup with dedicated multiple-channel micropump controller is available in LMSE.

 

Opis: Opis: Opis: Opis: Opis: DSCN7294Opis: Opis: Opis: Opis: Opis: C:\Users\pečar\Desktop\04072013165.jpg

 

Diffuser micropumps

Valves in check valves micropumps are sensitive to small particles and suffer from many problems such as wear and fatigue with long-term usage, induce difficulties such as clogging and sedimentation and have a rather complicated design. Therefore, they are not suitable for extreme miniaturization. To overcome these problems, the use of passive fixed valves as substitutes for active check valves is suggested. These pumps utilize the different pressure drop characteristics of flow through a nozzle and a diffuser to direct the flow in one preferential direction and hence cause a pumping action.

 

Opis: Opis: Opis: Opis: Opis: figOpis: Opis: Opis: Opis: Opis: DSCN6825

 

Strip-type microthrottle pump

A main challenge in development of micro pumps suitable for integration in lab-on-chip devices has been to produce a pumping mechanism that enables simultaneous and separate pumping of several liquids in parallel. Strip-type micropump with its rectangular actuator enables more efficient chip surface consumption (lab-on-chip: separate parallel pumping of media) compared to existing micropumps with circular actuators. Numerical simulation enabled virtual prototyping of the proposed device with detailed analysis of device operation at various frequencies, back-pressure, amplitude of the excitation signal as well as optimization of several design parameters such as optimal position and heights of the throttles.

 


 

Self-priming bubble tolerant microcylinder pump

The invention relates to a new type of micropumps which have the advantage of lower cost manufacturing, increased operating reliability and better performance than competing. Micropumps specialty is their ability of non-destructive pumping of biological substances which enables effective use in biomedical applications. Micropumps could be upgraded with advanced integration of other actuator, sensor and thermo components, thereby making it possible to create dedicated applications in biotechnology, biomedicine, chemistry, pharmacy, ecology, industry and more. In the Laboratory of Microsensor Structures and Electronics we have developed the entire process technology for micropumps production, so that it can be cost-effective to produce even a small series of dedicated products. In addition, we have developed electronics which enables proper device functioning in final application.

 

 

Features:

·       flow rate up to 2,4 ml/min

·       back pressure up to 0,4 bar

·       piezoelectric actuator

·       actuation amplitude: 50 to 300 V

·       biocompatible

·       size: 23x23x15 mm

·       weight: 2,4 g

·       materials: glass, PDMS

Applications:

·       microfluidic systems

·       drug delivery

·       biotechnology

·       chemical processing

·       medical equipment and devices

·       pharmaceutical and cosmetics

·       clinical and analytical lab

 

 

SQR excitation square-wave, 250 Vpp / 110 Hz

SQR excitation square-wave, 200 Vpp